A.1.3 Baseline Noise & Distortion Test - Time Domain Analysis
Connect the evaluation board’s analog input(s) to circuit ground.
Select the Time Domain window by clicking on the Time Domain tab at the top of the screen.
Click on the Acquire button. Within the Acquire sub-panel set Select Number of Samples to Capture to the
desired sample size. This example uses a sample size of 256 to display data with a high level of detail. To
examine a portion of the displayed data use the graphical zoom features located at the bottom right corner.