CRD5376
10 DS612RD2
Once the Setup panel is set, select Configure on the Digital Filter sub-panel.
After digital filter configuration is complete, click Capture on the Data Capture sub-panel.
Once the data record is collected, the Analysis panel is automatically displayed.
Select Noise FFT from the Test Select control to display the calculated noise statistics.
Verify the noise performance (S/N) is 124 dB or better.
1.4.2 Distortion Test
Set the controls on the Setup panel to match the picture:
Once the Setup panel is set, select Configure on the Digital Filter sub-panel.
After digital filter configuration is complete, click Capture on the Data Capture sub-panel.
Once the data record is collected, the Analysis panel is automatically displayed.
Select Signal FFT from the Test Select control to display the calculated noise statistics.
Verify the distortion performance (S/D) is 112 dB or better.